Ion Instrument Reprocessing

Manual Disinfection
Ion® Peripheral VISION PROBE
Endoscope Reprocessing Department wall chart steps are shown for the displayed device(s). Click on a step to view supplemental reprocessing information , or a video demonstration .
9
11
10
12
Rinse High Level Disinfection* Final Rinses(repeat 3x) Rinse under running cold water. Submerge in OPA, wipe off any bubbles from exterior, soak, and remove. Do not wipe pins. While submerged, wipe all surfaces, and soak in fresh critical water. Do not wipe pins.
13
Dry Final ERD Inspect Remove and dry exterior surfaces. Blow air as needed (optional). Do not wipe pins. Inspect entire device for soil under 2x magnification.
Dry Wipe exterior and tip with a fresh sterile, lint-free cloth soaked in alcohol. Blow air over connector and pins (optional).
1
2
3
Remove Accessory Unscrew, remove and dispose of the vision probe adapter. Inspect Inspect entire probe, connector, cable and tip for damage.
4
Prepare Solution Leak Test Prepare, run, confirm, and disconnect leak tester.
5
Soak in Cleaning Solution Submerge, wipe exterior surfaces, and then soak. Do not wipe pins.
6
7
8
Soak in Clean Water Rinse While submerged, wipe exterior surfaces, and then soak in clean, cold water. Do not wipe pins. Rinse under running cold water.
Prepare cleaning solution (pH 7-11). Follow manufacturer's concentration and temperature.
Store Store in a clean, dry place according to facility policy.
14

*Follow disinfectant manufacturer's instructions for soak time and temperature.

All steps reference: 555214-04 Rev. A

LEAK TEST PORT

Failure to secure the cap on the leak test port will result in device damage.

Leak test port

Secure cap on leak test port